RENT ROHDE & SCHWARZ ZV-Z229 Calibration Kit (Call for Availability)
Effective system data of R&S®ZV-Z229
The specified effective system data are established after performing a suitable system error calibration, e.g. TOSM, at a R&S® ZVA, R&S® ZVB, or R&S® ZVT vector network analyzer, using the characteristic data of the calibration kit, which are stored on a provided USB stick. This data is valid between +18 °C and +28 °C, at a measurement bandwidth of 10 Hz, and a nominal power of −10 dBm at the calibration ports. The calibration kit is fully functional down to 0 Hz, with effective system data as specified below, although the data is only verified for frequencies as stated.
Specifications for the ROHDE & SCHWARZ ZV-Z229 Calibration Kit Include:
Directivity |
10 MHz to 700 MHz | > 33 dB, typ. 36 dB |
700 MHz to 24 GHz | > 38 dB, typ. 42 dB | |
24 GHz to 40 GHz | > 33 dB, typ. 36 dB | |
Source match |
10 MHz to 700 MHz | > 30 dB, typ. 36 dB |
700 MHz to 24 GHz | > 36 dB, typ. 40 dB | |
24 GHz to 40 GHz | > 30 dB, typ. 36 dB | |
Reflection tracking |
10 MHz to 700 MHz | < 0.2 dB, typ. 0.1 dB |
700 MHz to 24 GHz | < 0.1 dB, typ. 0.05 dB | |
24 GHz to 40 GHz | < 0.2 dB, typ. 0.1 dB | |
Load match | 10 MHz to 700 MHz | > 33 dB, typ. 36 dB |
700 MHz to 24 GHz | > 38 dB, typ. 42 dB | |
24 GHz to 40 GHz | > 33 dB, typ. 36 dB | |
Transmission tracking | 10 MHz to 700 MHz | < 0.2 dB, typ. 0.1 dB |
700 MHz to 24 GHz | < 0.1 dB, typ. 0.04 dB | |
24 GHz to 40 GHz | < 0.2 dB, typ. 0.08 dB |